Steam Quality Detector SQD-A-4F

SHINVA · Engineering

Steam Quality Detector SQD-A-4F

Fully automatic SHINVA steam quality detector, model SQD-A-4F, developed for continuous online monitoring of three critical steam quality parameters: superheat, dryness fraction and non-condensable gas content. Compliant with EN 285, HTM 2010 and GB 8599 standards. With pure air-cooled design and lithium battery power, it requires no external water or power supply, making it ideal for on-site validation of autoclaves, sterilization systems and pharmaceutical plants.

Key specifications

  • Continuous online monitoring of superheat, dryness fraction and non-condensable gases
  • Compliant with EN 285, HTM 2010 and GB 8599
  • Startup and preheating time below 15 minutes
  • Testing cycle below 5 minutes
  • Measurement range — dryness fraction: 0.80 to 1
  • Measurement range — superheat: 0 to 50 °C
  • Measurement range — non-condensable gases: < 20%
  • Battery life of 6 hours or more
  • Simplified installation: only inlet hose connection required
  • Fully automatic design with one-touch start
  • Pure air-cooled design: no external cooling water required
  • Lithium battery powered: no external power source, safe and portable
  • User authority management, audit trail, data printing, storage and export

Applications

  • Pharmaceutical industry
  • Biological industry
  • Food industry
  • Hospitals and sterilization centers
  • Validation of autoclaves and pure steam systems
  • Process steam quality control
  • Regulatory compliance audits and inspections